VELOCITY #1 Preselecting fixed values for
VELOCITY #2 sound velocity

When calibrating the display range, you can select be-
tween two fixed sound velocity values in the coarse
mode.

In this function you determine which values are avail-
able for selection.

• You can set the sound velocity between 1000 m/s

and 9999 m/s (continuously adjustable).

Operation:

– Highlight the center function group in the left function

table with I.

– With M set the required values in VELOCITY #1

and VELOCITY #2.

H Note:

The preset default values here are as follows:

• 5918 m/s for longitudinal sound velocity

• 3353 m/s for transverse sound velocity

dB STEP Sets step range

You can set the gain with five different step ranges. In
this function you enter your own value for the adjust-
able step.

• You are able to enter a step range between 6.5 dB

und 24.0 dB in steps of 0.5 dB.

Operation:

– Highlight the center function group in the left function

table with I.

– With M set the required step range.

Configuring the USN 52R/USN 52L

Operation

Krautkramer USN 52R/USN 52L

Issue 05, 02/00

5-45

VELOCITY #1 Preselecting fixed values for
VELOCITY #2 sound velocity

When calibrating the display range, you can select be-
tween two fixed sound velocity values in the coarse
mode.

In this function you determine which values are avail-
able for selection.

• You can set the sound velocity between 1000 m/s

and 9999 m/s (continuously adjustable).

Operation:

– Highlight the center function group in the left function

table with I.

– With M set the required values in VELOCITY #1

and VELOCITY #2.

H Note:

The preset default values here are as follows:

• 5918 m/s for longitudinal sound velocity

• 3353 m/s for transverse sound velocity

dB STEP Sets step range

You can set the gain with five different step ranges. In
this function you enter your own value for the adjust-
able step.

• You are able to enter a step range between 6.5 dB

und 24.0 dB in steps of 0.5 dB.

Operation:

– Highlight the center function group in the left function

table with I.

– With M set the required step range.

Configuring the USN 52R/USN 52L

Operation

This manual is related to the following products: