Krautkramer USN 52R/USN 52L

5.7

Measuring thickness

USN 52R/USN 52L enable you to measure thickness
and distance in a simple way. The measured value is
directly displayed. One of the following two methods of
measurement is used, depending on the application:

• Measurement to the first echo – e.g. corrosion

measurement with a dual (T/R) probe

• Measurement of the multi-echo sequence with a

single- element probe on a test object which has a
good surface on both sides

Preparations

– Select probe

– Set MEASURE (refer to previous section)

– Select TOF (refer to Page 5-9)

– Calibrate A-scan display

– Calibrate the USN 52 R for the selected probe and

the measurement mode (refer to previous section)

Measuring the thickness

– Clean the surface of the test object from dirt and any

loose particles

– Apply couplant to the surface of the test object

– Using slight and even pressure, couple the probe

– Observe A-scan display and S=...”-value until the

measured value display is stable

– Maximize the echo amplitude

– Read measured value

H Note:

If you are operating with MULTECHO in MEASURE,
observe Gate b in order to evaluate the correct echo.

You can store the displayed measurement values in the
Data Logger or transfer them into another instrument
via the RS232 interface.

Operation

Measuring thickness

5-34

Issue 05, 02/00

Krautkramer USN 52R/USN 52L

5.7

Measuring thickness

USN 52R/USN 52L enable you to measure thickness
and distance in a simple way. The measured value is
directly displayed. One of the following two methods of
measurement is used, depending on the application:

• Measurement to the first echo – e.g. corrosion

measurement with a dual (T/R) probe

• Measurement of the multi-echo sequence with a

single- element probe on a test object which has a
good surface on both sides

Preparations

– Select probe

– Set MEASURE (refer to previous section)

– Select TOF (refer to Page 5-9)

– Calibrate A-scan display

– Calibrate the USN 52 R for the selected probe and

the measurement mode (refer to previous section)

Measuring the thickness

– Clean the surface of the test object from dirt and any

loose particles

– Apply couplant to the surface of the test object

– Using slight and even pressure, couple the probe

– Observe A-scan display and S=...”-value until the

measured value display is stable

– Maximize the echo amplitude

– Read measured value

H Note:

If you are operating with MULTECHO in MEASURE,
observe Gate b in order to evaluate the correct echo.

You can store the displayed measurement values in the
Data Logger or transfer them into another instrument
via the RS232 interface.

Operation

Measuring thickness