Issue 05, 02/00

5-23

Operation:

– In the function CAL press N and O simultaneously.

The function is now enabled.

REC 0 is displayed, indicating that no calibration

echo has been recordet yet.

– Set the gate onto the first calibration echo and enter

the sound path in S-REF using M.

– Save the first calibration echo using M in the CAL

function. On the display: REC 1.

– Set the gate onto the second calibration echo and

enter the sound path in S-REF using M.

– Save the second calibration echo using M in thte

CAL function. On the display: REC 2.

– Press O on CAL again to terminate the 2-step

calibration procedure.

Automatic calibration is carried out after the instrument
has calculated the entered parameters and made a
plausibility check.

H

Note:

To verify a correct instrument calibration, always check

the setting using the echo of a known distance, other
than those used for calibration.

Preparing the instrument for
digital measurement

USN 52R/USN 52L must be calibrated to the material to
be tested in order to give the correct measurement
readings.

H Note:

For all calibration procedures, use a calibration block
with a known thickness, made of the same material and
having the same temperature as your test object.

Preconditions:

To start with, calibrate the A-scan display as described
in the preceding section. Carry out the following basic
settings:

GATE LOGIC:
POSITIVE, NEGATIVE or MEASURE

MEASURE: 0 TO 1st – for measurement to the first
echo in the gate. By this, the thickness or the flaw posi-
tion can be measured.

USN 52R/USN 52L calibration

Operation

Krautkramer USN 52R/USN 52L

Issue 05, 02/00

5-23

Operation:

– In the function CAL press N and O simultaneously.

The function is now enabled.

REC 0 is displayed, indicating that no calibration

echo has been recordet yet.

– Set the gate onto the first calibration echo and enter

the sound path in S-REF using M.

– Save the first calibration echo using M in the CAL

function. On the display: REC 1.

– Set the gate onto the second calibration echo and

enter the sound path in S-REF using M.

– Save the second calibration echo using M in thte

CAL function. On the display: REC 2.

– Press O on CAL again to terminate the 2-step

calibration procedure.

Automatic calibration is carried out after the instrument
has calculated the entered parameters and made a
plausibility check.

H

Note:

To verify a correct instrument calibration, always check

the setting using the echo of a known distance, other
than those used for calibration.

Preparing the instrument for
digital measurement

USN 52R/USN 52L must be calibrated to the material to
be tested in order to give the correct measurement
readings.

H Note:

For all calibration procedures, use a calibration block
with a known thickness, made of the same material and
having the same temperature as your test object.

Preconditions:

To start with, calibrate the A-scan display as described
in the preceding section. Carry out the following basic
settings:

GATE LOGIC:
POSITIVE, NEGATIVE or MEASURE

MEASURE: 0 TO 1st – for measurement to the first
echo in the gate. By this, the thickness or the flaw posi-
tion can be measured.

USN 52R/USN 52L calibration

Operation