5-6
Issue 05, 02/00
Krautkramer USN 52R/USN 52L
Setting the receiver
All functions for adjustment of the receiver can be
found in the function group RCVR.
– With JC go to the function group RCVR.
FREQ. Frequency
With this function you can set the operating frequency.
You can choose between four settings:
• 0.4 - 10 MHz
= broadband amplifier
• 2 - 8 MHz
= 4MHz according to DIN 25 450
• 3 - 10 MHz
= high frequency range
• 0.3 - 4 MHz
= low frequency range
Operation:
– Select the required setting in FREQ. using M.
Operation
Adjustment of the USN 52R/USN 52L
5-6
Issue 05, 02/00
Krautkramer USN 52R/USN 52L
Setting the receiver
All functions for adjustment of the receiver can be
found in the function group RCVR.
– With JC go to the function group RCVR.
FREQ. Frequency
With this function you can set the operating frequency.
You can choose between four settings:
• 0.4 - 10 MHz
= broadband amplifier
• 2 - 8 MHz
= 4MHz according to DIN 25 450
• 3 - 10 MHz
= high frequency range
• 0.3 - 4 MHz
= low frequency range
Operation:
– Select the required setting in FREQ. using M.
Operation
Adjustment of the USN 52R/USN 52L