Krautkramer USN 52R/USN 52L

Issue 05, 02/00

5-31

5.6

Echo evaluation with
DAC/TCG

To be able to evaluate a flaw indication by means of the
DAC/TCG, certain requirements have to be met:

• The Distance-Amplitude Curve must already be

recorded.

• It

only applies to the same probe that was used for

recording the curve. Not even another probe of the
same type must be used!

• The curve and the DAC/TCG

only apply to the

material that corresponds with the material of the
reference block.

• All functions affecting the echo amplitude must be

set as they were during recording of the curve. This
especially applies to the following functions:
DAMPING, FREQ.,RECTIF., MTL VEL, PULSER
and REJECT.

TOF is set to PEAK.

AMPLITUDE is set to dB THRESHOLD.

• Choose the echo amplitude (as dB difference to the

curve) using the T V -keys for an enlarged
representation in the A-scan.

Echo evaluation with DAC/TCG

Operation

Krautkramer USN 52R/USN 52L

Issue 05, 02/00

5-31

5.6

Echo evaluation with
DAC/TCG

To be able to evaluate a flaw indication by means of the
DAC/TCG, certain requirements have to be met:

• The Distance-Amplitude Curve must already be

recorded.

• It

only applies to the same probe that was used for

recording the curve. Not even another probe of the
same type must be used!

• The curve and the DAC/TCG

only apply to the

material that corresponds with the material of the
reference block.

• All functions affecting the echo amplitude must be

set as they were during recording of the curve. This
especially applies to the following functions:
DAMPING, FREQ.,RECTIF., MTL VEL, PULSER
and REJECT.

TOF is set to PEAK.

AMPLITUDE is set to dB THRESHOLD.

• Choose the echo amplitude (as dB difference to the

curve) using the T V -keys for an enlarged
representation in the A-scan.

Echo evaluation with DAC/TCG

Operation