Operation:

The MULTECHO setting of the function MEASURE
(lower operating level) is a prerequisite for operation.

• Simultaneously press N and O in a-WIDTH or

a-THRESH in order to switch to the settings for
Gate b.

a-THRESH Determination of threshold

With this function you determine the threshold for the
selected gate. You can set a range from 10% to 90%
screen height in 1% steps.

Operation:

– Set the required threshold value in a-THRESH with

M.

The selected gate is displayed as a line.

If there is a violation of the set threshold (signals ex-
ceeding or falling below it), an alarm is activated.

The exact gate position is defined by two further
parameters:

• Gate start (a-START)

• Gate width (a-WIDTH)

a-START Start of the gate

In the function a-START you determine the start of
Gate a within a range from 0 to 5000 mm (continuous).

a-WIDTH Width of the gate

In the function a-WIDTH you determine the width within
a range from 1 to 5000 mm (continuous).

Operation:

– Set the gate position in a-START and a-WIDTH

with the assigned M keys.

Adjustment of the USN 52R/USN 52L

Operation

Krautkramer USN 52R/USN 52L

Issue 05, 02/00

5-11

Operation:

The MULTECHO setting of the function MEASURE
(lower operating level) is a prerequisite for operation.

• Simultaneously press N and O in a-WIDTH or

a-THRESH in order to switch to the settings for
Gate b.

a-THRESH Determination of threshold

With this function you determine the threshold for the
selected gate. You can set a range from 10% to 90%
screen height in 1% steps.

Operation:

– Set the required threshold value in a-THRESH with

M.

The selected gate is displayed as a line.

If there is a violation of the set threshold (signals ex-
ceeding or falling below it), an alarm is activated.

The exact gate position is defined by two further
parameters:

• Gate start (a-START)

• Gate width (a-WIDTH)

a-START Start of the gate

In the function a-START you determine the start of
Gate a within a range from 0 to 5000 mm (continuous).

a-WIDTH Width of the gate

In the function a-WIDTH you determine the width within
a range from 1 to 5000 mm (continuous).

Operation:

– Set the gate position in a-START and a-WIDTH

with the assigned M keys.

Adjustment of the USN 52R/USN 52L

Operation