beautypg.com

2 pr eset ting aut o matically – HEIDENHAIN TNC 320 (340 55x-03) Touch Probe Cycles User Manual

Page 78

background image

78

3.2 Pr

eset

ting aut

o

matically

8

Center in 1st axis

Q321 (absolute value): Center of

the pocket in the reference axis of the working plane.

8

Center in 2nd axis

Q322 (absolute value): Center of

the pocket in the minor axis of the working plane. If
you program Q322 = 0, the TNC aligns the hole center
to the positive Y axis. If you program Q322 not equal
to 0, then the TNC aligns the hole center to the
nominal position.

8

Nominal diameter

Q262: Approximate diameter of the

circular pocket (or hole). Enter a value that is more
likely to be too small than too large.

8

Starting angle

Q325 (absolute): Angle between the

reference axis of the working plane and the first touch
point.

8

Stepping angle

Q247 (incremental): Angle between

two measuring points. The algebraic sign of the

stepping angle determines the direction of rotation
(- = clockwise) in which the touch probe moves to the
next measuring point. If you wish to probe a circular
arc instead of a complete circle, then program the
stepping angle to be less than 90°.

X

Y

Q322

Q321

Q262

Q325

Q247

X

Z

Q261

Q260

MP6140

+

Q320

The smaller the angle, the less accurately the TNC can
calculate the datum. Minimum input value: 5°