The information for the comparison check is generated from several of the Report Properties

fields.

Report Field

Location in Tabs

Time

Chart Start Time

Comparison Test T/C

Spool Number

Middle Left of Comparison

Check

Comparison T/C

Number

Top Right of Comparison

Check

Comparison T/C Wire

Correction Factor

Top Right of Comparison

Check

Comparison

Instrument Correction

Factor

Thermocouples

Comparison T/C

Temperature Readings

Records the temperature

of the Comparison T/C

Adjacent T/C Number

for Comparison Check

Top Right of Comparison

Check

Adjacent T/C Wire

Correction

Thermocouples, click View

Offsets Used in Survey

button*

Adjacent Instrument

Correction Factor

Thermocouples, click View

Offsets Used in Survey

button

Adjacent T/C

Temperature Readings

for Comparison Check

Records the temperature

of Adjacent T/C

CC Difference

Difference Between

Comparison T/C and

Adjacent T/C temperatures

Accept/Reject Criteria =

Tolerance of Furnace

Set on Comparison Check

Page, Middle Right

Accept/Reject

Dependent on A/R Criteria

Table 2 - Location of Report Fields

*The offsets for the T/C Wire Correction can be adjusted solely for the purpose of the

comparison by checking the box on the Comparison Check page labeled Use Custom Wire

Corrections.

A header text and a body text can be put in the report. There are Preview Report and Print

Report buttons in the top left hand corner of the Comparison Check page.