beautypg.com

Calibration – Super Systems SDS 8120 User Manual

Page 31

background image

Super Systems Inc.

Page 31 of 104

Color Touch Screen Data Logger Operations Manual

Figure 38 - The About Screen

Calibration

Before using the SDS Data Logger in surveys, it is essential to ensure that the unit is properly

calibrated. This means that

raw

voltages

measured

by instruments connected to the data logger

are converted (scaled) to values that accurately reflect the

actual

voltages. The calibrations that

the user will typically need to perform (before using the data logger in surveys) are the Voltage

Calibration and Cold Junction Trim Calibration (also called CJ Trim). First, the Voltage

Calibration will be performed to ensure that the voltages measured by the leads connected to

each thermocouple are scaled correctly. Second, the CJ Trim will be performed to ensure that

the temperature measured at each data logger terminal (where the cold junction is created) is

correctly scaled. Both types of calibration require a sourcing device and thermocouple wire. The

sourcing device should be calibrated and certified to generate accurate voltages before these

calibrations are performed.

IMPORTANT:

Do not perform any of these calibrations without a sourcing device that is properly

connected and sourcing voltage into the proper inputs on the Color SDS Data Logger. Running a

calibration without a sourcing device that is providing voltage into the unit may require

additional calibration steps.

Voltage Calibration

The leads that connect to each input thermocouple measure the voltage generated by that

thermocouple. In most cases, the voltage that the data logger detects (the raw voltage) is

slightly different from the actual voltage generated by the thermocouple. When properly

performed, a Voltage Calibration (also called an Input Calibration) ensures that the SDS Data

This manual is related to the following products: