Super Systems Inc.

Page 39 of 104

Color Touch Screen Data Logger Operations Manual

Figure 49 - Reset Calibration Dates

IMPORTANT:

Avoid resetting calibration dates without first contacting Super Systems

Inc. Technical Support (

800-666-4330).

Set Nominals

Nominals are calibration constants used in the process of converting raw analog data to

voltage values. The Set Nominals menu provides the ability to reset calibration

constants for all inputs, or for a specified set of inputs, to default values.

A user may decide to use this feature if, for example, temperature readings from the

SDS are deemed to be outside a certain acceptable range and subsequent calibrations

are not successful. When the calibration constants are reset, the device is placed in an

uncalibrated state and will require recalibration. If recalibration does not result in

temperature readings that are within an acceptable range, it is likely that the SDS has

experienced hardware damage and will need to be shipped to Super Systems Inc. for

repair.


The Set Nominals screen is shown in Figure 50

.

To reset all nominals, simply select the

checkbox labeled “All” and then press the Set Nominals button.