Substitute leakage test, Flash test, Insulation test – Seaward Supernova Elite User Manual
Page 94
Insulation Test
84
Insulation Test
Test Voltage * ............... 500V d.c. or 250V d.c. nominal (0.5M
Ω load)
Short Circuit Current............................................. 2mA d.c. maximum
Range.......................................................................... 100k
Ω - 100 MΩ
Resolution.......................................................…..0.01 M
Ω (<100 MΩ)
Accuracy..................... +/- 5% of reading, +/- 2 digits (100k
Ω - 20MΩ)
Pass Levels ……...…0.5M
Ω, 1MΩ, 2MΩ, 4MΩ, 7MΩ, 10MΩ, 50MΩ
Substitute Leakage Test
Test Voltage * .....................................................................40V a.c. o/c
Display Range .......................................................... 0.1 - 20.0 mA a.c.
Resolution.........................................................................…....0.01mA
Accuracy..........................+/- 10% of reading, +/- 2 digits 1.00 - 20mA
Pass Levels …………………0.25mA, 0.5mA, 0.75mA, 2.5mA, 3.5mA
7.0mA, 9.9mA, 15.0mA
Flash Test
Test Voltage ....................….1.5kV a.c. (Class I), 3.0kV a.c. (Class II)
Display Range ............................................................ 0.1 - 2.5 mA a.c.
Resolution..........................................................................….0.01mA
Accuracy................................................ +/- 5% of reading, +/- 2 digits
Pass Levels.........................................1.0mA, 1.5mA, 2.0mA, 2.25mA
Insulation Test