Temperature monitoring
75
If the low voltage test fails, then either of the following warnings will
be displayed: -
--------------------WARNING--------------------
APPLIANCE MAY BE SHORT CIRCUIT
PRESS THE ENTER BUTTON IF SAFE
PRESS THE RED BUTTON TO ABORT
--------------------WARNING--------------------
FUSE MAY BE OPEN CIRCUIT OR
PRESS THE ENTER BUTTON IF SAFE
PRESS THE RED BUTTON TO ABORT
These messages mean that the EUT is switched off or the fuse may
be open circuit, or it maybe possible that the EUT will draw more than
18A and could damage the Tester. User discretion is required. If in
doubt abort the test and seek advice.
Temperature monitoring
The tester is provided with internal temperature monitors to ensure
sensitive components are not overheated. High rates of testing may
cause this situation, especially with long duration Earth Bond tests.
Setting Earth Bond test duration to 2 secs will increase operating
time. If the following message appears, leave the tester to cool
down before pressing Enter to continue testing.
WARNING
THE TEST HAS BEEN ABORTED BECAUSE THE APPLIANCE
TESTER BECAME TOO HOT. ALLOW THE UNIT TO COOL
BEFORE STARTING PRESS ENTER BUTTON TO CONTINUE