How to perform automatic tests, Use testcodes, Perform test – Seaward Supernova Elite User Manual
Page 33
How to perform Automatic Tests
23
Use Testcodes
This option bypasses the need to edit a specific test sequence,
allowing direct entry of a Testcode through the keyboard or a barcode
scanner.
Use the cursor keys to highlight the Testcode to be changed. Press
Enter to select that digit. Use the keyboard to change the value of
the digit. Press Enter to confirm the change. Press the Red button to
abort changes to the selected Testcode.
Use a barcode scanner to enter a Testcode
Use the cursor keys to highlight the PERFORM TEST option and
press Enter to commence the test sequence.
Press the Red button to return to the previous menu.
Perform Test
Use the cursor keys to highlight the Perform Test option and press
Enter to perform the test. Pressing the Red Button will return to the
previous menu.
The screen will change to the Meter Display (see separate section -
How to use the Meter Display for an explanation).
---------------ENTER TESTCODES-------------------------
OUTPUT LIMIT DUR NUM
SOCKET
230V
VISUAL
YES
1
EARTH BOND
4A 0.10
Ω 5 1
INSULATION
500V 0.50M
Ω 5 1
SUB LEAKAGE
SKIP
FLASH
SKIP
LEAKAGE
SKIP
TOUCH LEAKAGE
SKIP
POLARITY
SKIP
RCD
SKIP
TESTCODE
1:
I2F8008032
TESTCODE
2:
R222111111
PERFORM TEST
How to perform Automatic Tests