How to print Barcode Labels and Testcodes
55
Test Settings Editor
The TEST SETTINGS EDITOR allows the test sequence for the
selected test to be examined and changed if required.
Use the cursor keys to highlight a field to be changed and press
Enter to select the desired option.
Name
Each test sequence is given a name. This is to help the user when
selecting test sequences during automatic testing. When Name is
highlighted press the Enter button, then use the keyboard to enter a
new name for the Test Sequence (of up to 15 characters) and press
Enter to confirm.
Test Parameter Fields
By selecting one of the test parameter fields from the TEST
SETTINGS EDITOR, a list of options for that field will be displayed
INSULATION Test for example: -
---------------------TEST SETTINGS EDITOR-------------------
OUTPUT LIMIT DUR NUM
SOCKET 230V
VISUAL
YES
1
EARTH BOND 25A
∼ 0.10Ω 5s 1
INSULATION 500V
= 1.00MΩ 5s 1
SUB LEAKAGE
SKIP
FLASH
SKIP
LEAKAGE 230V
∼ 0.75mA 5s 1
TOUCH LEAKAGE
SKIP
POLARITY
SKIP
RCD
SKIP
TESTCODE 1: I1C700321
TESTCODE 2: S121111111
NAME: 230V HH / PORT 1
PRINT TESTCODES
OK