How to print barcode labels and testcodes, Earth bond limit calculator, Testcodes – Seaward Supernova Elite User Manual
Page 66
How to print Barcode Labels and Testcodes
56
-------------------INSULATION-----------------
O/P
LIMIT
DUR REP
250V SKIP
2s
0
500V 0.50MΩ
5s
1
1.00M
Ω
10s 2
2.00M
Ω
30s 4
4.00M
Ω
5
7.00M
Ω
10
10.00M
Ω
15
50.00M
Ω
20
25
30
UL
When you first enter this menu all of the previous values and one of
the Test Parameter columns are highlighted. Use the cursor left and
right keys to highlight the Test Parameter to be changed and then
use the cursor up and down keys to highlight the required setting.
Use this process for each Test Parameter.
Use the Enter button or OK Fast key to save the settings and return
to the TEST SETTINGS EDITOR. Press the Red button to abort
changes made to the settings and return to the TEST SETTINGS
EDITOR.
Earth Bond Limit Calculator
Earth Bond tests have an additional Fast Key; LIMIT CALC. Selecting
this enables the User to set an accurate Earth Bond Test limit by
specifying the cable parameters.
Testcodes
Selections made in the Test Parameter Fields will automatically set
the correct Testcode. Alternatively entering a Testcode will
automatically set the correct Test Parameters for that Testcode.
How to print Barcode Labels and Testcodes