Cementation with Multilink

®

Implant

For an optimum bond between the IPS e.max Press object and the Ti base, use the self-curing luting composite Multilink

®

Implant with light-curing option. Read the respective Instructions for Use for more detailed information.

Please observe the following procedure for cementation:
– Keep the cleaned and conditioned components that are to be luted (pressed objects, Ti base) at hand.
The subsequent cementation procedure must be carried out quickly and without interruption. The working

time of Multilink Implant is 90 (± 15) seconds at 23 °C (± 1 °C) or 73 °F (± 1.8 °F).

– As a general rule, attach a new mixing tip to the Multilink Implant syringe prior to each use.
– Apply Multilink Implant directly from the mixing tip in a thin layer on the bonding surface of the Ti base and the

bonding surface of the pressed object.

– Leave the mixing tip attached to the Multilink Implant syringe until the next application. As the material will cure in the

mixing tip, it will serve as a seal.

– Position the pressed object above the Ti base in such a way that the position markings are aligned.
– Use even and low pressure to join the parts and check the correct relative position of the components when they are in

their final position (transition Ti base/pressed object).

– Subsequently, tightly press the components together for 5 seconds.
– Carefully remove excess in the screw cavity, e.g. with Microbrush or brush, using a rotary movement.
– Remove excess at the transition to the Ti base carefully in its ductile state, e.g. with a foam pellet, while applying slight

pressure to hold the components in place.

– Apply a glycerine gel (e.g. Liquid Strip) to the cementation joint to prevent the formation of a inhibition layer.
– After that, completely polymerize the luting composite for 7 minutes in a light-polymerization device.
Important: Do not move the objects until the Multilink Implant material has completely cured and hold them

in place without allowing for any motion, e.g. with diamond-coated tweezers.

– After completion of the polymerization, rinse off the glycerine gel with water.
Smooth out and polish the cementation joint with rubber polishers.
– If there are any cement residues in the screw channel, remove them using suitable rotary instruments.
– Clean with a steam cleaner.

Keep the cleaned and conditioned components that are to be luted at hand.

Attach a new mixing tip to the Multilink Implant syringe prior to each use.