Material

Press­ceramic

IPS e.max Press are lithium disili-
cate glass-ceramic (LS

2

) ingots for 

the press technology. The industrial 
production process creates abso-
lutely homogeneous ingots in dif-
ferent translucency levels. For IPS e.max Press Abutment Solutions, 
ingots from the exisitng range of products are used.
The ingots feature a strength of 400 MPa and are thus the pressed 
ceramic ingots with the highest strength. They are pressed in Ivoclar 
Vivadent press furnaces to produce objects with outstanding accuracy 
of fit. The pressed, tooth-coloured, highly esthetic restorations are  
completed with IPS e.max Ceram.

Uses

Indications

–  Hybrid abutments for anterior and posterior single-tooth  

restorations

–  Hybrid abutment crowns for anterior and posterior restorations

Contraindications

–  Use of Ti bases which do not fulfil the geometry requirements.
–  Failure to observe the requirements stipulated by the implant 

 manufacturer for using the selected implant type (diameter and 
length of the implant must be approved for the respective position 
in the jaw by the implant manufacturer).

–  Bruxism
–  Failure to observe the permissible maximum and minimum ceramic 

wall thicknesses.

–  All uses not stated as indications are contraindicated. 

Important­processing­restrictions

Failure to observe the following restrictions may compromise the 
results achieved with IPS e.max Press:
–  If hybrid abutment crowns are fabricated, the opening of the 

screw channel must not be located in the area of contact points 
and areas with masticatory function. If this is not possible, a 
hybrid abutment with a separate crown would be preferred.

–  No extension units; only single-tooth restorations
–  Layering with a veneering ceramic other than IPS e.max Ceram
–  Pressing of two or more IPS e.max Press ingots in one investment 

ring

–  Pressing of IPS e.max Press in the IPS Investment System 300 g
–  Use of a luting composite other than Multilink

®

 Implant to lute 

IPS e.max Press to the Ti base

–  Temporary cementation of the crown on the hybrid abutment.
–  Failure to observe the manufacturer's instructions regarding the 

processing of the Ti base.

Side­effects

If the patient is known to be allergic to any of the components,  
IPS e.max Press Abutment Solutions should not be used.

   

CTE (100–400°C) [10

-6

/K] 

10.2

CTE (100-500°C) [10

-6

/K] 

10.5

Flexural strength (biaxial) [MPa]* 

400 

Fracture toughness [MPa m

0.5

] 

2.75

Modulus of elasticity [GPa] 

95

Vickers hardness [MPa] 

5800

Chem. solubility [µg/cm

2

]* 

40

Pressing temperature [°C] 

915–920

*according to ISO 6872

Ti­base

For IPS e.max Press Abutment Solutions, customary Ti bases made of 
titanium or titanium alloys are used. 
Please observe the instructions for use and processing of the manu-
facturer of the Ti bases used.