Hybrid abutment crown
Please observe the following procedure for the clinical try-in:
– Have the cleaned hybrid abutment crown (temporarily secured with Virtual Extra Light Body Fast Set) at hand.
– Remove the temporary restoration.
– Place the hybrid abutment crown intraorally onto the implant to check and possibly adjust the proximal
contacts. Note: No occlusal functional checks must be performed at this stage.
– Manually screw in the hybrid abutment crown with the matching screw.
– Check the geometry of the hybrid abutment crown (e.g. fit, gingival anaemia) with regard to the gingiva.
– Check the occlusion/articulation and make possibly required adjustments with suitable grinding instruments (see separate
IPS e.max Recommended grinding instruments for ceramics – use in the dental practice). If adjustments have been made
by grinding, conduct another polishing cycle or glaze firing.
– Carefully remove the hybrid abutment crown (including Ti base).
– Rinse the implant site e.g. with Cervitec Liquid (antibacterial mouth wash with chlorhexidine) to clean and disinfect it.
– Insert the temporary restoration.
Check the occlusion/articulation and use suitable grinding instruments to make
possibly required adjustments.
Manually screw in the hybrid abutment crown with the matching screw.
Place the hybrid abutment crown intraorally onto the implant to check and possibly
adjust the proximal contacts. Note: No occlusal functional checks must be
performed at this stage.
Carefully remove the hybrid abutment crown (including Ti base).
Check the geometry of the hybrid abutment crown (e.g. fit, gingival anaemia) with
regard to the gingiva.