Flowserve 3400IQ Digital Positioner User Manual
Page 36

Logix 3400IQ Digital Positioner FCD LGENIM3401-00 – 06/06
36
ExAMPLE:	The	Logix	3400IQ	digital	positioner	has	a	default	dead-band	configuration	of	20	percent.	
The	valve	has	a	4	inch	linear	stroke.	When	the	valve	first	powers	up,	the	command	signal	is	50	
percent.	The	unit	will	calculate	a	high	travel	threshold	of	70	percent	(50	percent	present	position	plus	
20	percent	dead-band)	and	a	low	travel	threshold	of	30	percent	(50	percent	present	position	minus	
20	percent	dead-band).	As	long	as	the	stem	position	remains	greater	than	30	percent	and	less	than	
70	percent,	no	additions	are	made	to	the	travel	accumulator.	Now,	assume	the	stem	position	moves	
to	80	percent	that	is	outside	the	present	dead-band.	The	Logix	3400IQ	digital	positioner	calculates	
the	stem	movement	and	adds	this	number	to	the	travel	accumulator.
80	percent	(present	position)	-	50	percent	(previous)	=		
30	percent	movement	x	4-inch	stroke	=	1.2	inches
So,	1.2	inches	is	added	to	the	travel	accumulator.	New	dead-band	thresholds	of	100	percent	(80	
percent	present	position	plus	20	percent	dead-band)	and	60	percent	(80	percent	present	position	
minus	20	percent	dead-band)	are	calculated.	This	process	continues	as	the	stem	position	moves	
throughout	its	stroke	range.
7.14.5 Cycle Counter
The	cycle	counter	is	another	means	of	monitoring	valve	travel.	Unlike	the	travel	accumulator,	the	
stem	position	must	do	two	things	to	count	as	a	cycle:	exceed	the	cycle	counter	dead-band	and	
change	direction.	A	cycle	counter	limit	can	also	be	written	into	the	positioner.	If	this	limit	is	exceeded,	
the	yellow	LED	will	blink.
7.14.6 Position Deviation
If	the	stem	position	differs	from	the	control	command	by	a	certain	amount	for	a	given	length	of	time,	
the	yellow	LED	will	blink	to	signify	excess	deviation.	The	trip	point	and	settling	times	are	set	from	the	
transducer	function	block.
7.14.7 Advanced Features
NOTE:	These	features	are	contained	in	the	transducer	function	block.	Refer	to	the	Reference	Manual	
for	a	more	detailed	explanation.
7.14.8 Standard vs. Advanced Diagnostics
Advanced	diagnostics	models	add	top,	bottom,	and	supply	pressure	sensors.	This	allows	for	more	
diagnostic	calculations	such	as	loss	of	pressure,	advanced	signatures,	and	troubleshooting.
7.14.9 Temperature and Pressure Units
The	desired	temperature	and	pressure	units	can	be	set	during	configuration.	Once	set,	all	readings	
will	be	displayed	in	the	desired	units.
