10.2.3 Output Calibration

The AC voltage output calibration is performed by adjusting the output up or down using a
calibration coefficient. The coefficient ranges from -127 (adjust upward) to + 127 (adjust
downward). An external traceable reference AC voltmeter should be used for this purpose.

CALibrate[:SOURce]:VOLTage

This command will set the calibration coefficient for the AC full scale output voltage.

Command Syntax

CALibrate[:SOURce]:VOLTage

Parameters

(a value between -127 and +127)

Examples

CAL:VOLT -2

Query Syntax

CALibrate:VOLTage?

Returned Parameters

(value range -127 to +127)

Related Commands

CAL:STAT