Example: Individual Phase Loss Test
The following command sequence will drop the output of phase A only for 10 cycles at 400
Hz.

/* Set up sequence
VOLT 115;FREQ 400

Set output voltage to 115 V at 400 Hz (all phases)

INST:COUP 0

Uncouple output phases

INST:NSEL 1

Select phase A

OUTP 1

Close output relay

OUTP:STAR:STAT 1

Enable start phase mode

OUTP:STAR:PHAS 90

Set start phase angle to 90 degrees on phase A

OUTP:DROP 0.025

Drop for 25 ms (10 cycles @ 400 Hz)

Output drop status query

The status of a drop in progress can be queried using the query format of the OUTP:DROP
command. This query will return a indicating if the output time specified has expired
(0) or the drop is still in progress (1).

Command Syntax

OUTPut:DROP

Parameters

| none

*RST Value

none

Examples

OUTP:DROP 0.001

Query Syntax

OUTP:DROP?

Returned Parameters

0 | 1

Related Commands

OUTP:STAR:STAT

OUTP:STAR:PHAS

INST:COUP

INST:NSEL