10.2.2 Measurement Calibration

All measurement calibrations are performed by adjusting the measurement reading up or
down using a calibration coefficient. The coefficient value ranges from -127 (adjust full-scale
downward, zero upward) to + 127 (adjust full-scale upward, zero downward). When used in
combination with the MEAS commands, an automated measurement calibration procedure
can be implemented by adjusting the coefficient and reading the new measurement value
iteratively until the delta between an external reference measurement device and the AC
source internal measurement is within the specified accuracy.

CALibrate:MEASure:CURRent[:FSCale]

This command affects the calibration of the AC current measurement at full scale.

Command Syntax

CALibrate:MEASure:CURRent[:FSCale]

Parameters

(value range -127 to +127)

Examples

CAL:MEAS:CURR -23

Query Syntax

CALibrate:MEASure:CURRent?

Returned Parameters

(value range -127 to +127)

Related Commands

CALibrate:MEASure:VOLTage

CALibrate:MEASure:CURRent:ZERO

This command affects the calibration of the AC current measurement zero offset.

Command Syntax

CALibrate:MEASure:CURRent:ZERO

Parameters

(value range -127 to +127)

Examples

CAL:MEAS:CURR:ZERO +12

Query Syntax

CALibrate:MEASure:CURRent:ZERO?

Returned Parameters

(value range -127 to +127)

Related Commands

CALibrate:MEASure:CURRent

CALibrate:MEASure:PCURrent[:FSCale]

This command affects the calibration of the AC peak current measurement at full scale.

Command Syntax

CALibrate:MEASure:PCURrent[:FSCale]

Parameters

(value range -127 to +127)

Examples

CAL:MEAS:PCUR -23

Query Syntax

CALibrate:MEASure:PCURrent?

Returned Parameters

(value range -127 to +127)

Related Commands

CALibrate:MEASure:VOLTage

CALibrate:MEASure:PCURrent:ZERO

This command affects the calibration of the AC current measurement zero offset.

Command Syntax

CALibrate:MEASure:PCURrent:ZERO

Parameters

(value range -127 to +127)

Examples

CAL:MEAS:PCUR:ZERO +12

Query Syntax

CALibrate:MEASure:PCURrent:ZERO?

Returned Parameters

(value range -127 to +127)

Related Commands

CALibrate:MEASure:CURRent