4.6 Remote Inhibit [Option]
A Remote Inhibit BNC input is located on the rear panel of the 2003RP if the -OP1 option
was specified at the time of purchase. This input may be used to disable the AC source
output using an external control signal.
Default mode of operation
If the Remote Inhibit (RI) input is left unconnected, the output of the AC source can be turned
on using the OUTPUT button on the front panel or the appropriate bus command. To disable
the output relay, a logic low level or a contact closure needs to be present across the
terminals of the RI BNC connector.
Typical use of the RI input is in rack systems where the RI input is wired to a switch that
detects removal of the rear panel. This input may also be used to simultaneously shut down
multiple AC power sources.
Note that the RI input has an internal pull-up resistor to TTL logic high. When the RI input
cable is disconnected, the output will recover and the output relay will be closed
automatically.
- CW-M (48 pages)
- CW-M Corrected Table 4-2 in (1 page)
- CW-P (62 pages)
- Lx Series (205 pages)
- CW Series Programming Manual (25 pages)
- Ls Series II Programming Manual (242 pages)
- Compact i/iX Series (157 pages)
- Compact IX 2253 (157 pages)
- Compact i/iX Series Software Manual (203 pages)
- ASD Series Quick Start (5 pages)
- ASD Series (120 pages)
- i-iX Series II Programming Manual (226 pages)
- DLM 600W Series Programming Manual (24 pages)
- M131 Programming Manual (99 pages)
- DLM Series (74 pages)
- DLM 600W Series (82 pages)
- BPS Series (153 pages)
- DLM600 Series (16 pages)
- DCS-E 1.2kW Series (65 pages)
- DLM-E 4kW Series Programming Manual (32 pages)
- M136 (8 pages)
- DCS-E 3kW Series (94 pages)
- CTS 3.0 (166 pages)
- CSW Series (174 pages)
- 2001RP (131 pages)
- MX CTSH (151 pages)
- MXCTSL Administrator Manual (27 pages)
- MX CTSL (157 pages)
- RS Series (228 pages)
- MX Series Installation Manual (35 pages)
- Ls AC source (2 pages)
- MX15 Series (184 pages)
- Ls Series II (226 pages)
- Lx Series Driver Manual (275 pages)
- MX Series Rev: AY (257 pages)
- iX Series (341 pages)
- i-iX Series II (258 pages)
- GUPS 2400A-108 (36 pages)
- HPD Series (58 pages)
- HPD Series Operation Manual (41 pages)
- HPD Series GPIB-Multichannel (134 pages)
- PLA-PLW Programming Manual (74 pages)
- ReFlex Mating Connnectors for Controller (3 pages)
- LPDC-16V (4 pages)