6

Calibration

Routine calibration should be performed every 12 months. Non-routine calibration is only
required if a related assembly is replaced or if the periodic calibration is unsuccessful.
Calibration can be performed through the front panel by putting the 2003RP in Calibration
mode (CAL mode) or over the RS232C or IEEE-488 bus if the interface option is installed.
For bus calibration, the PGUI32 program may be used. Refer to the PGUI32 help file
supplied with the program for instructions on calibration over the remote control bus.

All routine calibrations can be performed without removing the top cover (closed case
calibration). For the

–AV option, program 115V or 230V for 135V or 270V respectively.

6.1 Calibration Equipment

Digital Multimeter (DMM):

Two (2) Fluke 8506A or 8508A

100 milliohm Current Shunt:

Isotek Model RUG-Z-R100-0.1 or equivalent

Load Bank:

Various power load resistors will be needed

Model

Range

Current

max.

Load

2003RP

135 V

5.0 A

27

270 V

2.5 A

108

2003RP-AV

115 V

5.9 A

20

230 V

4

2.9 A

78

2003RP-HV

156 V

4.3 A

36

312 V

2.16 A

144

Table 6-1: Load Resistors and Current

4

Some 2003RP-AV may be factory configured for single voltage range only. In this case, the 230 V

range is not available.