2 mounting the wastewater can, Caution – BINDER KMF 115 User Manual
Page 74
![background image](https://www.manualsdir.com/files/767134/content/doc074.png)
KMF (E5.3) 07/2014
page 74/98
(3) Hose connections
Plug the freshwater hose into the hose connection (21) above the freshwater can and secure it with
a hose clamp. You can use a part of the standard supplied water hose.
Screw the hose nozzle (brass) to the free edge of the hose and screw it directly onto the freshwater
connection “IN” (12) at the rear of the unit.
When the freshwater can is empty, within 60 seconds the alarm message HUMID ALARM will be dis-
played on the controller, the buzzer sounds (chap. 11), and the humidity system turns off.
To guarantee humidification during 24 hours even at high humidity set-points with manual wa-
ter supply, we recommend filling the freshwater can (option) at the end of each day.
14.7.2 Mounting the wastewater can
(1) Fixing (if required)
Hang the can with its holding device on its 4 carriers at the free space next to the freshwater can.
(2) Hose connections
Plug the wastewater hose to the hose connection (23) of the can and secure it with a hose clamp.
You can use a part of the standard supplied water hose.
Wastewater hose connection (23)
Figure 25: Wastewater can (option)
Plug the free hose edge to the wastewater connection “OUT” (11) at the rear of the unit and secure
it with a hose clamp.
You can remove the wastewater can with its holding device for emptying (disconnect the hose first before
emptying).
CAUTION
Overflow of the wastewater can.
Damage to the surrounding.
Empty the wastewater can in a timely manner before it is full.
Bringing a source of humidity into the inner chamber may increase wastewater production.
Regularly check the filling level of the wastewater can.