beautypg.com

6 intended use, Danger – BINDER KMF 115 User Manual

Page 15

background image

KMF (E5.3) 07/2014

page 15/98

1.6 Intended use

Constant climate chambers series KMF are suitable for exact conditioning of harmless materials. A mix-

ture of any component of the charging material with air must NOT be explosive. The operating tempera-

ture must lie below the flash point or below the sublimation point of the charging material.
Other applications are not approved.
Constant climate chambers series KMF are not classified as medical devices as defined by the

Medical Device Directive 93/42/EEC.

Following the instructions in this operating manual and conducting regular maintenance work

(chap. 15) are part of the intended use.

DANGER

Explosion hazard.
Danger of death.

Do NOT introduce any substance combustible or explosive at working temperature into

the constant climate chamber.

NO explosive dust or air-solvent mixture in the inner chamber.

The charging material shall not contain any corrosive ingredients that may damage the ma-

chine components made of stainless steel, aluminum, and copper. Such ingredients include in

particular acids and halides. Any corrosive damage caused by such ingredients is excluded

from liability by BINDER GmbH.

WARNING: If customer should use a BINDER chamber running in non-supervised continuous

operation, we strongly recommend in case of inclusion of irrecoverable specimen or samples

to split such specimen or samples and store them in at least two chambers, if this is feasible.

This manual is related to the following products: