Start up, Ole_link18, Ole_link19 – BINDER KMF 115 User Manual
Page 32

KMF (E5.3) 07/2014
page 32/98
5. Start up
•
After connecting the supply lines (chap. 4), turn on the unit via the main power switch (2).
•
Open the water-tap for freshwater supply. Alternatively, fill the freshwater can (option, chap. 14.7).
•
Turn on the humidifying and dehumidifying system with switch (3) (humidity switch ON/OFF).
After the first turning on of the chamber or after an interruption of the power supply the relative humidity
will increase after a delay of about 20 minutes. During this period, the relative humidity can drop consider-
ably.
Warming chambers may release odors in the first few days after commissioning. This is not a quality de-
fect. To reduce odors quickly we recommend heating up the chamber to its nominal temperature for one
day and in a well-ventilated location.
WARNING: If customer should use a BINDER chamber running in non-supervised continuous
operation, we strongly recommend in case of inclusion of irrecoverable specimen or samples
to split such specimen or samples and store them in at least two chambers, if this is feasible.
5.1 Function overview of the MB1 display program controller
W
X
CONFIG
HAND
VIEW->
EXIT
W
08:43:55 20.12.05
TEMP
°C
25.0
36.8
HUMID
%rH
50.0
58.6
Pilot lamp: Ready for operation
EXIT button (to exit a menu point)
AUTOMATIC button (to start a previously entered program)
ENTER button (to confirm a selection)
Navigation buttons (functions are assigned by the menu)
Figure 12: MB1 temperature and humidity program controller
The MB1 2-channel program controller controls following values inside the constant climate chamber:
Channel 1: Temperature in °C (range without humidity: -10 °C / 14 °F to 100 °C / 212 °F)
Channel 2: Relative humidity in % r.H. (range 10 % r.H. to 90 % r.H.)
You can enter the desired set point values in Manual Mode or Program Mode (chap. 5.2) in the display
controller. For the control range of temperature and relative humidity, see chap. 12).