3 menu settings in the “user-settings” menu – BINDER KMF 115 User Manual

Page 37

background image

KMF (E5.3) 07/2014

page 37/98

6.3 Menu settings in the “User-settings” menu

User-settings

Instrument data

Contrast

Displ. Power down

Contin. operation

User Level

Safety control. Act + 39.8 °C

Safety control. Set. + 42.0 °C

29

Instrument data

Instrument Name
Enter an individual name of the constant climate chamber.
Address
Enter a controller address (1 to 30) for operation with the communication

software APT-COM™.
All other entries are relevant only for service purposes.

Contrast

(no function)

Displ. power down

Switch off event
Do not change the entry “Wait. Period”.
Waiting period
You can enter a delay time after which the display, following manual activa-

tion, will automatically be turned off. This happens when the moment is out-

side the operation time defined in menu ”Contin. operation”.

Contin. operation

Enter an operation time to determine the period of display activity. Outside the

defined time, the display is automatically turned off. Pressing down any key

will reactivate the display. After the time set in menu “Displ. power down“, the

display will turn off again when the actual time is not within the operation time

fixed in menu “Cont. operation“.

User Level

Toggle here to the display menu “User Level” (chap. 6.4) by entering a pass-

word. Factory default setting for this password is +00001. You can change the

password (“user code”) in the menu “User Level”.

Safety control.Set

The setting of the tolerance limit of the safety controller (chap. 10.2) is dis-

played. You cannot change it in this view.

Safety control.Act

The measuring result of the safety controller (chap. 10.2) is displayed. The

safety controller compares the value measured by a second independent

temperature sensor to the entered tolerance limit.