12. Humidity system

The humidity system is turned on with the humidity switch (3) located in the right lateral control panel.
The constant climate chamber KMF is equipped with a capacitive humidity sensor. This results in a control

accuracy of up to ± 3 % r.H. of the set point. The temperature-humidity diagrams (Figure 18) show the

possible working ranges for humidity.

The preset temperature and humidity values should be situated within the optimum range

(hatched range in Figure 18).Only within this area will the unit not be exposed to excessive

moisture due to condensation.
In the short-term set points outside the optimum range can also be targeted. The control accu-

racies of ± 3 % r.H., however, cannot be guaranteed in this case.

0

10

20

30

40

50

60

70

80

90

100

10

20

30

40

50

60

70

80

90

100

% r.H.

°C

A

B

C

D

Figure 18: Temperature-Humidity diagram KMF


Range A: Ensured condensation free range
Range B: Temporary usable operation (up to 24 h)
Range C: In this range, condensation in the inner chamber is possible
Range D: In this range, deviations of the technical data are possible

Heat emission of electrical devices connected inside the chamber may modify the temperature

and humidity range.

Entry of the humidity set-point 0 % in defined program sections permits completely turning off

humidity in Program Mode and thus attaining faster temperature changes.

The constant climate chamber KMF is equipped with a door heating system to prevent condensa-

tion in the door area.
If the set points for temperature or humidity are outside the optimum range, condensation can arise in the

door area.