With set-point type “

Limit”, the user shall adapt the safety controller (chap. 10.2) or the tem-

perature safety device class 3.3 (option, chap. 10.3) to the highest temperature set-point value

of the program actually used. Check the safety device for each temperature program and

change it if necessary. Set the safety controller set-point or the set-point of the temperature

safety device class 3.3 (option) by approx. 2 °C to 5 °C above the controller temperature set-

point.
Recommended setting: Set-point type “

Offset” with set-point 2 °C.

Performance after completing the program:
The controller changes to Idle Mode. Heating, refrigeration, and humidification are inactive; the chamber

approximates ambient temperature. The fan turns at 50% rate (factory setting).

9.3 Entry of humidity values

08:43:59 15.12.13

Prog. Select.

Prog 1 PROG 01

Prog 2 PROG 02

Prog 3 PROG 03

Prog 4 PROG 04

Prog 5 PROG 05

Prog 6 PROG 06

Prog 7 PROG 07

Prog 8 PROG 08

Prog 9 PROG 09

Prog 10 PROG 10

Prog 11 PROG 11

Prog 1 PROG 01

TP- Program 2

TP- Program 3

TP- Program 1

Select the second subroutine “

TP-Program 2” and confirm by pressing “ENTER”. A program table will

appear which is initially empty until you enter the temperature values. You can now enter the humidity

program.
Humidity at the beginning

of the program section



Fan speed in %

(no function)

Program No.

Subprogram TP-Program No. 2

08:43:55 15.10.13

2

Total number of program

sections


Parameter set (preselected)
Tolerance band limits humidity

(maximum and minimum

humidity)

Operation lines

(no function)

Duration of program

section

Number of start section in case of

repeat cycles

Number of repeats in case of

repeat cycles


Proceeding further is equivalent to the temperature value entry described in chap. 9.2.
Time course of the subroutines:
When starting the overall program, both subroutines (TP-Program 1 and TP-Program 2) run off synchro-

nously. They should be of the same duration because each of the subroutines becomes inactive after run-

off (i.e., no heating or refrigeration and 50% fan speed after ZP 1 is completed, no humidification after ZP

2 is completed). When the complete program is finished, the controller changes to Idle Mode. Tempera-

ture and humidity proceed towards ambient values.
Performance after completing the program:
The controller changes to Idle Mode. Heating, refrigeration, and humidification are inactive; the chamber

approximates ambient temperature. The fan turns at 50% rate (factory setting).