Program entry as “set-point ramp” (example of a temperature program)

W/°C

t/min.

01

02

03

04

0

20

40

60

80

100

30

120

180

380

05

Tolerance

maximum

Tolerance

maximum

Program table corresponding to the diagram above:

Program

section

Set-point

temp.

Fan Section time Operation

lines

Target

section

No. of

cycles

Min.

tolerance

Max.

tolerance

01

40

100 %

00:30:00

Off

1

0

-1999

+9999

02

60

100 %

01:30:00

Off

1

0

-5

+5

03

90

100 %

01:00:00

Off

1

0

-2

+2

04

90

100 %

03:20:00

Off

1

0

-1999

+9999

05

20

100 %

00:00:01

Off

1

0

-1999

+9999

Now enter the values of the above program table into one of the 25 program places of the MB1 controller: