4.2 Wastewater connection
Fasten the wastewater hose to the wastewater connection “OUT” (11) on the rear of the unit (olive ∅ 14
mm). Observe the following points:
•
You can use a part of the supplied water hose as a drainage hose. In case another hose is used, it has
to be permanently resistant against at least 95 °C / 203 °F.
•
Mount the wastewater hose with a maximum positive inclination of 1 m and a maximum total length of
3 m.
•
Protect both ends of the drainage hose with two of the four supplied hose clamps.
Waste water is collected in an internal can with a volume of approx. 0.5 liters. It is pumped off
only when required, thus there is no continuous waste water flow.