11. Notification and alarm functions

11.1 Notification and alarm system overview (auto diagnosis system)

The unit provides notification and alarm functions, which indicate messages in up to three steps:
1. Visual indications of notifying or error messages are blue notes on the display of the MB1 controller.
2. Visual indications of alarm messages are red notes with an alarm bell symbol. After a delay time, some

notes change their color from blue to red.

In addition, there is an audible alert, if you did not deactivate the buzzer in the “User level” menu (chap.

6.4).


Notifying sequence

1

2

Event

Note

(blue field)

Alarm

(red field)

Error of the humidity system. Press RESET key

for further information.

HUMID SYSTEM

immediately

HUMID SYSTEM

after 1 min.

Notification or error of the humidity system.
For detailed information see chap. 11.2.

HUMID SYSTEM

(notification flashes)

Temperature deviation of more than +/- 2 °C of

the entered set-point

TEMP RANGE

immediately

TEMP RANGE

after 16 min.

Humidity deviation of more than +/- 5 % r.H. of

the entered set-point

HUMID RANGE

immediately

HUMID RANGE

after 16 min.

Exceeded temperature limit of the safety con-

troller

TEMPERATURE LIMIT

immediately

Humidity system turned off with switch (3)

HUMID OFF

immediately

Door open

DOOR OPEN

immediately

DOOR OPEN

after 5 min.

With option temperature safety device class 3.3 (chap. 10.3):
Exceeding the maximum or minimum tempera-

ture

TEMP ALARM

immediately

With option keyboard locking (BINDER Individual, chap. 14.5):

Keyboard locked

KEY LOCK

immediately

The indicated intervals refer to the time after occurrence of the error or notified condition.