beautypg.com

Options, 2 interface rs 422 (option), 3 data logger kits (option) – BINDER KMF 115 User Manual

Page 70

background image

KMF (E5.3) 07/2014

page 70/98

14. Options

14.1 Communication software APT-COM™ 3 DataControlSystem (option)

The constant climate chamber is regularly equipped with an Ethernet interface (6) that can connect the

BINDER communication software APT-COM™ 3 DataControlSystem. The actual temperature and humidi-

ty values are given at adjustable intervals. Programming can be performed graphically via PC. Up to 30

chambers with RS 422 interface can be cross-linked. The MAC Address is indicated below the Ethernet

interface. For further information, please refer to the operating manual of the BINDER communication

software APT-COM™ 3.
The additional RS422 interface (7) is only used for service purposes. Do NOT connect it to any network.

The interface is labeled accordingly.

14.2 Interface RS 422 (option)

With this option, the chamber is equipped with a serial interface RS 422 (replacing the Ethernet interface)

that can connect the BINDER communication software APT-COM™ 3 DataControlSystem. The actual

temperature and humidity values are given at adjustable intervals. For further information, please refer to

the operating manual of the BINDER communication software APT-COM™ 3.
Pin allocation of the RS 422 interface: Pin 2:

RxD (+)

Pin 3:

TxD (+)

Pin 4:

RxD (-)

Pin 5:

TxD (-)

Pin 7:

Ground

14.3 Data logger kits (option)

BINDER Data Logger Kits offer an independent long-term measuring system for temperature and humidi-

ty, available for different temperature ranges. According to the selected kit, the Data Logger can measure

and record also the ambient temperature and humidity values via a second multi-function sensor.
BINDER Data Loggers are equipped with a keyboard and a large LCD display, alarm functions and a real-

time function. Measurement data are recorded in the Data Logger and can be read out after the meas-

urement via the RS232 interface of the Data Logger. It offers a programmable measuring interval and

permits storing up to 64000 measuring values. Reading out is done with the Data Logger evaluation soft-

ware. You can give out a combined alarm and status protocol directly to a serial printer.
Data Logger Kit TH 100: Multi-function sensor for chamber temperature and humidity: Temperature

range -40 °C / -40 °F up to +100 °C / 212 °F , humidity range 0% r.H. up to 100% r.H.
Data Logger Kit TH 100/70: Multi-function sensor for chamber temperature and humidity: Temperature

range -40 °C / -40 °F up to +100 °C / 212 °F, humidity range 0% r.H. up to 100% r.H. Multi-function sensor

for ambient temperature and humidity: Temperature range -40 °C / -40 °F up to 70 °C / 158 °F, humidity

range 0% r.H. up to 100% r.H.

For detailed information on installation and operation of the BINDER Data Logger, please

refer to the mounting instructions Art. No. 7001-0204 and to the original user manual of the

manufacturer, supplied with the data logger.