10.3 Temperature safety device class 3.3 (DIN 12880) (option)
With the option over/under temperature protective device (temperature safety device class 3.3 acc. to DIN
12880) the unit is equipped with two additional safety devices (class 3.1 and class 3.2). The combination
of the safety devices is regarded as a safety device class 3.3.
The temperature safety device, class 3.3, serves to protect the constant climate chamber, its environment
and the contents from exceeding the maximum permissible temperature. Please observe the guideline
BGI/GUV-I on safe working in laboratories (formerly BGR/GUV-R 120 or ZH 1/119 laboratory guidelines
issued by the employers’ liability insurance association) (for Germany).
With
safety device class 3.1 a maximum value for the temperature is set that the unit will not exceed due
to the regulatory function of the safety device class 3.1. This protection against excessively high tempera-
tures can, for example, serve to protect the constant climate chamber, its environment and the material
under treatment from excess temperatures.
With
safety device class 3.2 a minimum value for the temperature is set that the unit will not fall below
due to the regulatory function of the safety device class 3.2. This protection against excessively low tem-
peratures can, for example, serve to protect sensitive loads from under cooling.
Both safety devices are functionally and electrically independent of the temperature control system. If an
error occurs, they perform regulatory function.
Safety devices class 3.1 (8) and class 3.2 (9) are located in the left lateral control panel.
With option temperature safety device class 3.3, the safety controller (chap. 10.2) must be set
to limit 100 °C / 212 °F.
(9)
(8)
Excess
temperature
Set-point
Insufficient
temperature
W
X
CONFIG
HAND
VIEW->
EXIT
W
08:43:55 20.09.02
TEMP
°C
40.0
36.8
HUMID
%rH
65.0
58.6
(1)
Figure 15: Temperature safety device class 3.3