8 Maintenance and service
60
Version 08/2009, Rev. 2.1 of 02/12/2014 • sb
Translation of the original operating manual
• Remove the cable cover at the bottom of the unit.
6
Cable cover
Fig. 53: Removing the cable cover at the bottom of the unit
The cables of the electronic temperature measurement system of the shaft
can be seen now.
• Unscrew the two terminal screws of the electronic temperature
measurement system of the shaft.
7
Terminal screws
Fig. 54: Removing the terminal screws of the electronic temperature measurement system
• Turn the rotational vacuum concentrator back to its normal position
(with a sufficient number of people helping you), if it was turned upside
down.