Rotational Vacuum Concentrator RVC 2-33 IR

6 Operation

40

Version 08/2009, Rev. 2.1 of 02/12/2014 • sb

Translation of the original operating manual

6.5.2 Mode

The mode selection can be activated by pressing the left-hand function key
"mode" in the active values window. The following phases are available:
• Press the left-hand function key "mode". The menu "start with phase…"

appears (see figure).

• Select the desired menu item with the up and down keys.
• Press the right-hand function key "enter" to confirm.

Fig. 22: Selecting the mode


Opening the lid
The lid can only be opened if the rotor is at a standstill. It is not possible to
open the lid during an evaporation process.

DANGER

During the evaporation process, the housing, the lid and the interior of the
rotational vacuum concentrator can reach surface temperatures of more
than +50°C.
Risk of burns!


Warm-up
Before starting an evaporation process, the vacuum pump and the cooling
trap should warm-up until they reach their respective operating
temperatures.
For this purpose, the warm-up phase can be started. The rotor can
optionally be installed. The warm-up time and the set value for heating can
be selected (see chapter 6.5.3.5 - "Options").
During the warm-up:
• The rotor chamber is preheated - with or without rotor - when the lid is

closed.

• The vacuum pump is switched on.
• The cooling trap is switched on if it is connected via the control system

with the rotational vacuum concentrator (see chapter 5.6 - "Connection
of a vacuum pump and/or a cooling trap").If there is no connection, the
cooling trap must be switched on separately.

When the preset warm-up time is over, the control unit displays the
following message:

Fig. 23: Message after the warm-up