8.1.2 Rotor chamber
If the rotor chamber is contaminated or if glass is broken, the rotor
chamber, the rotor shaft and the rotor must be cleaned immediately in order
to prevent corrosion.
Cleaning of the rotor chamber
• Open the lid.
• Disconnect the power cord from the wall outlet or instrument receptacle.
DANGER
Hot surface!
Let the rotor chamber cool down to room temperature!
1
Rotor chamber
2
Electronic
temperature
measurement system
of the shaft
3
Cap
4
Rotor shaft
5
Magnetic coupling on
rotor side
Fig. 52: Rotor chamber with rotor shaft incl. magnet coupling
• To get access to the bottom of the unit, turn the unit upside down or
place the unit between two trestles or two tables with a sufficient
number of people helping you.
CAUTION
The rotational vacuum concentrator weighs approx. 49 kg!