5 set-up and connection, 1 installation site, 2 power supply – Martin Christ RVC 2-33 CDplus User Manual
Page 26
Rotational Vacuum Concentrator RVC 2-33 IR
5 Set-up and connection
26
Version 08/2009, Rev. 2.1 of 02/12/2014 • sb
Translation of the original operating manual
Pos: 86 / 200 Christ /37 0 RVC-BA (STANDARDM ODULE)/ 050 Au fstellu ng u nd Ans chluss/ 050 Aufst ellung un d Anschl uss= === == == === == == === == == == == @ 25\ mod _14 049 850 8287 8_6 8.d ocx @ 186 981 @ 1 @ 1
5 Set-up and connection
Pos: 87 / 200 Christ /37 0 RVC-BA (STANDARDM ODULE)/ 050 Au fstellu ng u nd Ans chluss/ 050 -00 10 A ufstello rt, Eins atzo rt @ 25\ mod _14 0498 508 370 7_6 8.do cx @ 1 869 95 @ 2 @ 1
5.1 Installation site
Operate the RVC only in closed and dry rooms.
• The table must be stable and have a solid, even tabletop.
• Ensure sufficient ventilation. Do not place any paper, cloth, or similar
material behind or under the unit, since otherwise the air circulation will
be impaired.
• Keep a safety distance of at least 30 cm from the wall so that the vents
in the unit remain fully effective.
• The ambient temperature must be in the range of +10°C to +25°C.
• Do not subject the RVC to thermal stress, e.g. by positioning it near
heat generators.
• Avoid direct sunlight (UV radiation).
Pos: 88 / 010 Unive rsalm odul e/Le erzeil e @ 0\mo d_1 202 116 244 500 _0.d ocx @ 114 @ @ 1
Pos: 89 / 200 Christ /37 0 RVC-BA (STANDARDM ODULE)/ 050 Au fstellu ng u nd Ans chluss/ 050 -00 20 E ner gieve rsor gun g-- -- --- -- --- --- -- --- -- --- -- --- --- -- --- - @ 25\ mod _140 498 508 458 5_6 8.doc x @ 1 870 09 @ 2 @ 1
5.2 Power supply
Pos: 90 / 200 Christ /37 1 RVC-BA (PROJEKTE) /RVC 2-3 3 IR/ 050 Au fstellu ng u nd Ans chluss/ 050 -00 20- 001 0 Ansc hlussa rt RVC 2 -33 IR @ 31\ mod _140 748 406 576 8_6 8.doc x @ 2 090 14 @ 3 @ 1
5.2.1 Connection
DANGER
The operating voltage on the name plate must correspond to the local
supply voltage!
CHRIST rotational vacuum concentrators are units of safety class I.Units of
the type RVC 2-33 IR have a three-wire power cord with an IEC C13
connector (see chapter 10 - "Technical data").
An equipotential bonding screw is located on the back (see chapter 2.1.1 -
"Functional and operating elements"). This equipotential bonding screw can
be used to perform an earth conductor check.
Pos: 91 / 010 Unive rsalm odul e/Le erzeil e @ 0\mo d_1 202 116 244 500 _0.d ocx @ 114 @ @ 1
Pos: 92 / 200 Christ /37 0 RVC-BA (STANDARDM ODULE)/ 050 Au fstellu ng u nd Ans chluss/ 050 -00 20- 002 0 Sich eru nge n ba useits @ 25 \mo d_1 404 9850 854 66_ 68.d ocx @ 187 023 @ 3 @ 1
5.2.2 Customer-provided fuses
Typically, the RVC must be protected with 16 Amp G fuses that are to be
provided by the customer.
Pos: 93 / 010 Unive rsalm odul e/Seite nwechs el @ 0 \mo d_1 202 116 244 312_ 0.d ocx @ 105 @ @ 1