7 Malfunctions and error correction
56
Version 08/2009, Rev. 2.1 of 02/12/2014 • sb
Translation of the original operating manual
Pos: 167 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /07 0 Stör ung en u nd Fehl ersuc he/ 070 -00 10- 005 0 Kleinfl ansch ver bindu nge n @ 25\ mod_ 140 498 519 019 9_6 8.docx @ 1 874 85 @ 3 @ 1
7.1.5 Small flange connections
Leakages are often due to improper small flange connections between the
various components and hose connections or to leakages in the valves.
• Loosen the connection and place the centring ring (with sealing ring
inside) in a centred manner between the flange connections.
• Seal the connection with the clamping ring or clamping chain by
tightening the wing nut.
• Ensure that the centring ring neither slips out of place nor gets jammed.
Fig. 48: Small flange and centring ring
Fig. 49: Small flange with centring ring
and small flange
Fig. 50: Attaching the clamping ring
Fig. 51: Tightened clamping ring