Appendix C: RapidVap Specifications
Product Service Domestic 1-800-522-7658, International 816-333-8811
RapidVap N
2
/48 Models
The following tables show the approximate time to
evaporate various solvents in 12mm and 20mm sample
tubes when the sample block is fully loaded. Other sample
blocks are available for different size tubes. If they are
used or if a different number of samples is run at one time,
the evaporation times can be estimated from these charts or
determined experimentally. Evaporation rates may vary
from tube to tube.
Solvent:
Methylene
Chloride
Toluene Acetonitrile Methanol
Hexane/
Ethyl Acetate
75:25
Hexane/
Ethyl Acetate
50:50
Block Preheat
& Block Set
Point Temp:
40°C
80°C
75°C
40°C
40°C
40°C
Starting Volume
2 ml
6
8
8
18
5
5
Evaporation Time RapidVap N
2
/48 Models
(12 mm Sample Tubes) (Time in Minutes)
Number of samples: 48
Vortex speed: 100%
Nitrogen pressure: 8 psi
N
- RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910014 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910013 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910012 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910011 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910010 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910001 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910000 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900013 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900012 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900011 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900010 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900003 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900002 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900001 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900000