beautypg.com

Labconco RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910015 User Manual

Page 75

background image

71

Appendix C: RapidVap Specifications

Product Service Domestic 1-800-522-7658, International 816-333-8811







Solvent:

Methylene

Chloride

Ethyl

Acetate

Hexanes

Acetone

Toluene

Water

Block

Preheat &
Block Set
Point Temp:



40°C



75°C



70°C



55°C



100°C



95°C

Glassware

Ending

Volume

0.5 ml Stem

0.5

Dry

4

10

4
9

3
7

4
9

5

10

42
78

1.5 ml Stem

1.5
1.0
0.5

Dry

4
9

17
31

4
7

13
24

2
5
8

16

3
7

12
22

5
9

15
26

23
43
75

120

2.0 ml Stem

2.0
1.0

Dry

3

17
64

3

11
34

2
8

26

3

15
47

4

14
50

28
80

201

3.0 ml Stem

3.0
2.0
1.0

Dry

4

20
57

128

4

12
34
81

3

10
25
52

3

16
50

108

5

17
40
89

34
76

170
339







Table C

Evaporation Rate in Cool-Zone Stem

(600 ml Sample Tubes) (Time in Minutes)

Number of samples: 8
Vortex speed: 70%
Nitrogen pressure: 10 psi
Sample tubes: Covered or not covered with caps

Glassware Used:
0.5 ml (PN 79259-00) .77 in (1.9 cm) Long Stem
1.5 ml (PN 79139-00) 1.37 in (3.5 cm) Long Stem
2.0 ml (PN 79137-00) 1.62 in (4.1 cm) Long Stem
3.0 ml (PN 79135-00) 2.25 in (5.7 cm) Long Stem

N