Chapter 7: Troubleshooting
Product Service Domestic 1-800-522-7658, International 816-333-8811
PROBLEM
CAUSE
CORRECTIVE ACTION
No gas flow
(RapidVap N
2
)
Gas depleted
Install new supply of gas.
Sample position not
activated
Activate position per page 30.
Nozzle clogged
Unplug nozzle with fine wire.
Valve inoperable
Contact Labconco
Gas flows
continuously
Valve inoperable
Contact Labconco
Control inoperable
Contact Labconco
This manual is related to the following products:
- RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910014 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910013 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910012 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910011 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910010 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910001 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910000 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900013 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900012 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900011 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900010 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900003 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900002 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900001 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900000