Appendix C: RapidVap Specifications

Product Service Domestic 1-800-522-7658, International 816-333-8811

Evaporation Rates

RapidVap Vacuum Models


The following tables show the approximate time to
evaporate various solvents in 75mm and 16mm sample
tubes when the sample block is fully loaded. Other sample
blocks are available for different size tubes. If they are
used or if a different number of samples is run at one time,
the evaporation times can be estimated from these charts or
determined experimentally.

Evaporation Time – 75 mm dia. Sample Tube

(Time in Minutes)



Solvent:

Methylene

Chloride

Toluene

Acetonitrile

Water

Block Preheat &
Block Set Point
Temperature:

50°C

80°C

75°C

80°C

Vacuum:

250 mbar

133 mbar

167 mbar

133 mbar

Starting Volume

300ml 62

90

102

402

250 56

72

84

345

200 49

66

71

289

150 42

57

60

239

100 36

33

49

189

50 22

40

37

95


Number of samples: 8
Vortex speed: 28%
Vacuum pump: Labconco model – 79249-00

Rated free air flow – 88 L/min

Max. attainable vacuum – 31 mbar