Chapter 4: Using Your RapidVap
Product Service Domestic 1-800-522-7658, International 816-333-8811
4. Find “Constant b” for the solvent on the top of the
grid. Follow that line until it intersects with the line
as drawn between the two points. This point
represents the boiling point at the given pressure
and the temperature is read off the left or right side
of the grid (0-250°C).
To determine the operating pressure necessary to
achieve a given boiling point:
1. Find the boiling point at normal atmospheric
pressure (1 Bar) on the left side of the graph.
2. Find the “Constant b” and follow this line to the
desired boiling point on the grid (0-250°C).
3. Draw a straight line linking these two points and
extrapolate on the right hand scale to find the
operating pressure.
- RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910014 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910013 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910012 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910011 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910010 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910001 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7910000 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900013 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900012 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900011 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900010 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900003 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900002 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900001 RapidVap Vacuum, N2 and N2/48 Evaporation Systems 7900000