Methods – Hanna Instruments HI 903 User Manual

Page 84

background image

5-20

METHODS

5.5.14.1 Start Mode

The user can select the starting mode for the titration. In

Cautious

mode, the titrant dosing

begins with the minimum dose in order to prevent over-titration. In

Normal

mode, the

titrant dosing begins with the median value between the minimum and maximum (i.e.
minimum dose 5

µ

L, maximum dose 25

µ

L, first dose will be 15

µ

L).

5.5.14.2 Standby Mode

When enabling this option the titrator will automatically revert to Standby mode after a
titration is completed.
See also the

Standby Duration

option.