Methods – Hanna Instruments HI 903 User Manual
Page 79

5-15
METHODS
5.5.10.5 External Solvent Size (
External Dissolution/Extraction Determination Modes Only).
Enter the mass of the solvent used for external dissolution or extraction of the sample.
Weigh the solvent after determining the solvent water content but before adding sample to
the solvent.
5.5.10.6 External Solvent Conc. (
External Dissolution/Extraction Determination Modes Only).
5.5.10.7 Extracted Sample Size
(
External Extraction Determination Mode Only).
5.5.10.8 Dissoluted Sample Size
(External Dissolution Determination Mode Only).
5.5.11
Standard (
Titrant Standardization mode only)
This screen allows the user to define a list of Karl Fischer standards and customize related
parameters.
Using the
and
keys, highlight the standard from the existing list and press
to
choose it.
Press
if you want to create and add a new standard to the Karl Fischer standard data base.
Press
if you want to remove a Karl Fischer standard from the pre-defined list.
Press
if you want to edit the Karl Fischer standard parameters.