Method ID: HI8301EN

Moisture Determination in Solvent with 5 mg/mL Titrant (One-Comp.)

for external dissolution or extraction

Titrator Application Methods: General Methods - r. 0.4

Description:

Method for the determination of moisture in
extraction/dissolution solvent using 5 mg/mL One-
Component Titrant. The results are expressed in %
mass and should be less than 0.1%.

Electrode:

HI 76320 Double Platinum Pin Electrode


Reagents:

5 mg/mL One-Component KF Volumetric Titrant

Dry Methanol


Other Accessories:

A clean, dry 1-mL syringe

A clean, dry 22-gauge

, 6” non-coring septum

penetration needle

GL 45-Thread Solvent Bottle


Titration Procedure:

Setup titrator according to the instruction manual.

Press “Select Method” from the main screen. Use
t

he arrow keys to highlight ‘HI8301EN Solvent w/

5mg/mL 1-comp.

’ and press “Select”.

Install a 5-mL burette filled with 5 mg/mL One-

Component KF Volumetric Titrant and verify that
no air bubbles are present in the burette or tubing.
If necessary prime until all air has been removed
completely.
For the determination of the exact concentration of
the titrant, follow HI8001EN 5.0 mg/mL Titrant
Standardization using Liquid Water Standard.

Connect the solvent bottle top assembly to the

bottle of methanol according to the manual.

Prepare the titration vessel according to the

manual. Dispense enough solvent from the solvent
bottle t

o fill the vessel to the “min” line (about 50

mL).

Press “Start/Stop” to pre-titrate the solvent and the
titration vessel moisture. Allow the background
drift rate to stabilize before proceeding to the next
step.

Stop stirring the solvent in the extraction/

dissolution bottle.

Fill

the

syringe

and

needle

with

the

extraction/dissolution solvent.

Weigh the syringe, needle and solvent.

Press “Start Analysis”. You will be prompted to
enter the sample size.

Dispense 0.7500 g to 1.0000 g of solvent into the

titration vessel through the septum using the
needle. Pay attention not to get any solvent on
the electrode or beaker walls. If necessary swirl
the titration vessel by hand.

Clear the needle of residual solvent by intaking a

small volume of air from the titration vessel. If a
“hanging drop” of solvent is seen on the end of the

needle, dip the end of the needle briefly in the
titration solvent.

Remove the needle from the titration vessel and

weigh the syringe again in order to determine the
added sample mass (by difference of the two
measurements.)

Use the numeric keypad to enter the exact weight

and press “Enter” to start the analysis.

At the end of the titration the “Sample Analysis
Result

” screen is displayed. The results are

expressed in % mass of water. Record this value
as the “External Solvent Concentration”.

Method Parameters:

Name:

Solvent w/ 5mg/mL 1-comp.

Method Revision:

1.0

Type:

Sample Analysis

Predispensing Amount:

None

Pre-Analysis Stir Time:

0 Sec

Stirring Speed:

900 RPM

Stirbar Type:

Medium

Drift Entry:

Automatic

Solvent:

Methanol

Sample Parameters:

Sample Determ.:

Normal

Sample Name:

Solvent

Sample Type:

Mass

Sample Size:

1.0000 g

Titrant:

Composite 5

Titrant Type:

One Component

Nominal Titrant Conc.:

5.0000 mg/mL

Std. Titrant Conc.:

5.0000 mg/mL

Date/Time:

Jan 1, 2011 12:00

Titrant Age Reminder:

2d:00h:00m

Control Parameters:

Start Mode:

Cautious

Standby Mode:

Enabled

Standby Duration:

720 minutes

Imposed Current:

20 uA

Minimum Dose:

0.2500 uL

Maximum Dose:

5.0000 uL

Timed Increment:

1 second

End Point Value:

180.0 mV

Signal Averaging:

3 Readings

Flow Rate:

10.0 mL/min

Termination Parameters:

Maximum Duration:

600 sec

Maximum Titrant Volume:

5.0000 mL

Term. Criterion:

Relative Drift

Relative Drift:

10.0 ug/min

Result Unit:

%

Calculations:

Titrant units:

mg/mL

Titrant volume consumed:

V (mL)

Final results units:

% Mass

Titrant concentration:

5.0000 mg/mL

Sample mass:

1.0000 g