Quick start guide, How to perform the first titration – Hanna Instruments HI 903 User Manual
Page 11
11
QUICK START GUIDE
HOW TO PERFORM THE FIRST TITRATION
Method Selection
For this analysis we will use the HI8301EN Solvent with 5mg/ml 1-component
Titrant standard method
.
To select this method:
• Press
from the
Idle
screen. Use the
and
keys to highlight the
HI8301EN Solvent with 5mg/ml 1-component Titrant method.
• Press
.
After accomplishing these operations, the method’s name will be displayed on the top line of
the
Idle
screen.
Setting Method Parameters
To display the method parameters, press
. The
View/Modify Method
screen will be
displayed.
Only certain parameters from the standard methods can be changed.
For this titration, only the KF titrant concentration value and the size of the solvent sample
need to be entered as in the screen shown below.
To accomplish this:
• Highlight
Titrant
option from the
View/Modify Method
screen and then press
.
• The
Karl Fischer Titrants
screen will be displayed.
• Press
.
• Highlight
“Standardized Titrant Concentration”
and
press
.
• Input the correct value, then press
.
• Press
three times to return to the
Idle
screen.
Setup Titration Report
Users can select the information that is stored for each titration that is performed.
To obtain proper information at the end of the titration, perform the following operations:
• From the main screen, press
and the
Data Parameters
screen will be
displayed.