5-4

METHODS

From PC:
Unnecessary Standard Methods can be removed from the titrator using the HI 900 PC
application (see Section

4.9, USB Link with PC

).

5.2

Standard Methods

The standard methods were developed for the most common types of analysis. Also, the
standard methods can be used as a model to create new user methods.
Only specific method parameters can be modified by the user (see Section

5.5,

Method

Options

section).

5.2.1

Upgrading Standard Methods

To upgrade the titrator with new standard methods, follow the steps below:

From USB Storage Device:

Insert the USB storage device into the USB port, located on the left side of the titrator.

Access the

General Options

screen.

Using the

and

keys, highlight the

Restore Files from USB Storage Device

option and choose

.

Using the

and

keys, navigate through file types to find “standard method

files”. The list with available standard methods on the storage device will be displayed.

Press the

or

key to upgrade the titrator with the standard methods.

Press

to return to

General Options

screen.

Note: See section 4.8 Restore Files from USB Storage Device.

From PC:
You can upgrade the titrator with standard methods from a PC using the HI 900 PC application
(see Section

4.9, USB Link with PC

).

5.2.2

Deleting Standard Methods

Unnecessary standard methods can be removed from titrator by following the procedure
below:

From

General Options

screen:

Access the

General Options

screen.

Using the

and

keys, highlight the

Save Files to USB Storage Device

option and

press

;

Using the

and

keys, navigate through the file types to find “standard method

files”. The available standard methods will be displayed.

Press the

or

keys to remove unnecessary standard methods.

Press

to return to the

General Options

screen.