Methods – Hanna Instruments HI 903 User Manual
Page 68

5-4
METHODS
From PC:
Unnecessary Standard Methods can be removed from the titrator using the HI 900 PC
application (see Section
4.9, USB Link with PC
).
5.2
Standard Methods
The standard methods were developed for the most common types of analysis. Also, the
standard methods can be used as a model to create new user methods.
Only specific method parameters can be modified by the user (see Section
5.5,
Method
Options
section).
5.2.1
Upgrading Standard Methods
To upgrade the titrator with new standard methods, follow the steps below:
From USB Storage Device:
•
Insert the USB storage device into the USB port, located on the left side of the titrator.
•
Access the
General Options
screen.
•
Using the
and
keys, highlight the
Restore Files from USB Storage Device
option and choose
.
•
Using the
and
keys, navigate through file types to find “standard method
files”. The list with available standard methods on the storage device will be displayed.
•
Press the
or
key to upgrade the titrator with the standard methods.
•
Press
to return to
General Options
screen.
Note: See section 4.8 Restore Files from USB Storage Device.
From PC:
You can upgrade the titrator with standard methods from a PC using the HI 900 PC application
(see Section
4.9, USB Link with PC
).
5.2.2
Deleting Standard Methods
Unnecessary standard methods can be removed from titrator by following the procedure
below:
From
General Options
screen:
•
Access the
General Options
screen.
•
Using the
and
keys, highlight the
Save Files to USB Storage Device
option and
press
;
•
Using the
and
keys, navigate through the file types to find “standard method
files”. The available standard methods will be displayed.
•
Press the
or
keys to remove unnecessary standard methods.
•
Press
to return to the
General Options
screen.